Measured effects reported in the paper
| What was changed | Where | Measured | Effect |
|---|---|---|---|
| add contrastive learning to 1-D CNN | CNN · VTaC real-time | Challenge Score: 76.17 -> 79.07 | Improved |
| add contrastive learning to FCN | FCN · VTaC real-time | Challenge Score: 80.08 -> 78.41 | Hurt |
| add contrastive learning to FCN | FCN · VTaC retrospective | Challenge Score: 75.84 -> 78.95 | Improved |
| add contrastive learning to CNN | CNN · VTaC retrospective | Challenge Score: 76.94 -> 78.85 | Improved |
| add 5 seconds of post-alarm data (retrospective vs real-time) | CNN · VTaC | Challenge Score: 76.17 -> 76.94 | No clear effect |
| add 5 seconds of post-alarm data (retrospective vs real-time) | FCN+CL · VTaC | Challenge Score: 78.41 -> 78.95 | Improved |
| add 5 seconds of post-alarm data (retrospective vs real-time) | FCN · VTaC | Challenge Score: 80.08 -> 75.84 | Hurt |
| replace rule-based method by FCN | Rule-based (Plesinger) · VTaC real-time | Challenge Score: 67.32 -> 80.08 | Improved |
| replace CNN by FCN as feature extractor | CNN · VTaC real-time | AUC: 0.936 -> 0.949 | Improved |
| replace supervised CNN by Transformer encoder | CNN · VTaC real-time | Challenge Score: 76.17 -> 62.73 | Hurt |
| use generative anomaly detection (Diffusion+CL) instead of supervised FCN | FCN · VTaC real-time | Challenge Score: 80.08 -> 52.51 | Hurt |
| use autoencoder reconstruction loss (SAE) instead of plain MLP | MLP · VTaC real-time | Challenge Score: 45.58 -> 68.77 | Improved |
Closest papers by meaning
- Yet Another ICU Benchmark: A Flexible Multi-Center Framework for Clinical ML (ICLR 2024 · similarity 0.82)
- Reading Your Heart: Learning ECG Words and Sentences via Pre-training ECG Language Model (ICLR 2025 · similarity 0.81)
- GEM: Empowering MLLM for Grounded ECG Understanding with Time Series and Images (NeurIPS 2025 · similarity 0.81)
- Parallel-Learning of Invariant and Tempo-variant Attributes of Single-Lead Cardiac Signals: PLITA (AAAI 2025 · similarity 0.81)
- From Token to Rhythm: A Multi-Scale Approach for ECG-Language Pretraining (ICML 2025 · similarity 0.81)
- ECG-Reasoning-Benchmark: A Benchmark for Evaluating Clinical Reasoning Capabilities in ECG Interpretation (NeurIPS 2026 · similarity 0.80)
Writing a paper in this area? scime draft is a free LaTeX editor with AI included, real-time co-authors and a check of your references and venue rules before you submit. It also imports OpenAI Prism projects.