AI Paper Atlas

Research map / 3D vision / Depth and stereo

X-ray Computed Tomography Through Scatter

Adam Geva, Yoav Y. Schechner, Yonatan Chernyak, Rajiv Gupta

ECCV 2018 · 3D vision › Depth and stereo · approach: Non-line-of-sight imaging · 16 citations (OpenAlex)

Read the paperCheck your idea against 137,217 papers

AI-extracted. The measured effects below were read by AI from the paper's full text (no summary yet). Numbers and claims can be wrong or missing context, so check the paper before you cite or build on it.

Measured effects reported in the paper

What was changedWhereMeasuredEffect
replace linear tomography with scattering CT (no ASG)XCAT hand reconstruction · XCAT hand, total volumeepsilon error: 80% -> 30%Improved
replace linear tomography with scattering CT (no ASG)XCAT hand reconstruction · XCAT hand, Z slice 40epsilon error: 76% -> 28%Improved
replace linear tomography with scattering CT (no ASG)XCAT hand reconstruction · XCAT hand, Y slice 50epsilon error: 24% -> 18%Improved
replace linear tomography with scattering CT (no ASG)XCAT hand reconstruction · XCAT hand, total volumemass error delta_mass: 70% -> 1%Improved
initialize multi-scatter analysis with single-scatter approximation (ablation)Scattering CT · XCAT thigh (9x10^9 photons)CNR: 0.76 -> 1.02Improved
use scattering CT without ASG instead of linear tomography with ASGXCAT thigh simulation · XCAT thigh quality vs dosedose for same epsilon error: x5 more dose for the same epsilon errorImproved